66--moveable table and hood for robotic instrument - small business set-aside (OB1669552) (2004-12-31)

NAICS Code: 334516 This is a combined synopsis/solicitation for commercial
items prepared in accordance with the format in FAR Subpart 12.6, as
supplemented with additional information included in this notice. This
announcement constitutes the only solicitation; quotations are being requested
and a written solicitation will not be issued. Solicitation number
RFQ-OH-05-00028 is issued as a Request for Quotation. This acquisition is a 100%
small business set-aside. The NAICS code is 334516. The enclosure shall be a
Class I or II type cabinet as described in NSF/ANSI 49-2004, section 3.4 that
is exhausted through a HEPA filter in a one pass format into the room exhaust
ductwork located directly overhead. It shall feature a front sash or door that
allows easy viewing of and access to all interior areas, an internal 110V
electrical outlet with attached, 6 ft. power cord for connection to a wall
outlet and internal fluorescent lighting. Internal UV lighting is also desirable
but not essential. The enclosure and table shall have sufficient internal
dimensions and strength to enclose and support with minimal vibration a Qiagen,
BioRobot 8000 (width: 58 in., height: 31 in., depth: 32 in., weight: 300 lb).
All materials shall withstand normal wear, corrosive action of gasses and
fluids, cleaning compounds and decontaminating agents and procedures. Materials
shall be structurally sound, dimensionally stable and moisture resistant.
Exposed interior surfaces shall be smooth and abrasion and corrosion resistant
or shall be rendered corrosion resistant with non-toxic material that resists
crazing, cracking and chipping. Other interior and exterior surfaces shall be
smooth and abrasion and corrosion resistant or shall be rendered corrosion
resistant with non-toxic material that resists crazing, cracking and chipping.
Windows shall be optically clear and not adversely affected by accepted cleaning
methods and decontaminating agents. Protective coatings shall be resistant to
prolonged contact to liquids, cleaning compounds and procedures. Specifically
the protective coatings shall be resistant to the following chemicals: 1 N
hydrochloric acid; 1 N sodium hydroxide; 1% quaternary ammonium hydroxide; 5%
formaldehyde; 5000 ppm hypochlorite; 2% iodophore, 5% phenol and 70% ethyl
alcohol. When exposed to these chemicals there shall be no visible effect on
the finish other than a slight change in gloss, discoloration and/or temporary
softening of the finish with no loss of adhesion or film protection. The
enclosure/table shall be designed and fabricated to function properly and
operate in a safe manner, minimize contamination, provide personnel protection
and be capable of being cleaned and decontaminated. Exposed burrs and sharp
edges (including but not limited to sheet metal screws) shall be eliminated from
those surfaces of the cabinet that are subject to normal operation, field
certification and maintenance. The enclosure shall be mounted on adjustable
legs, or other acceptable means to assure a minimum of 4 in of unobstructed
clearance beneath the unit. Legs and feet shall be sufficiently rigid to provide
support with a minimum of cross bracing and shall be of simple design. Doors and
covers shall fit properly and close completely. HEPA filters shall conform to
the materials, construction and aerosol efficiency requirements of
IEST-RP-CC-001 for type C filters. Filter media shall conform to MIL-F-51079B.
Filter units of sizes not covered by these specifications shall be of the same
type, materials and construction that have been qualified for the filter
manufacturer. In addition, HEPA filters shall be scan tested for a leakage not
to exceed 0.01% when tested in accordance with section A.3 of NSF/ANSI 49-2004
(see below). The enclosure shall be designed to provide accessibility for
filter installation, testing and sealing. The filters shall be mounted to
prevent air bypass of the filters and shall be protected with a filter guard to
prevent damage to the filter and blockage of exhaust air. Fan motors shall be
sized to operate at a static pressure sufficient to meet the performance
requirements of the enclosure (smoke test, see below). A thermal protector
shall be provided and it shall not trip at 115% of the rated voltage under
maximum load and ambient temperature conditions. The motor shall be rated for
continuous operation and shall be variable speed with controls that are located
outside the work area. Motors and lights shall be protected from the
receptacles. Circuit overload protection, conforming to the National Electrical
Code shall be provided for the motor and all receptacles. Flexible power cords
for single phase power shall be 3 wire, with the ground wire connected to the
frame, unless otherwise specified and sized in accordance with the National
Electrical Code for the specified load(s). A wiring diagram showing connection
of all electrical components as well as a statement providing starting current,
running power and circuit requirements shall be provided with the installation
instructions. The external dimensions of the enclosure shall not exceed 82 in.
width x 37 in. depth x 8 ft. height when fully assembled. It shall be possible
with minimal adjustments (e.g. of the legs) to move the enclosure through a 7 ft
height x 37 in. width door frame. The table shall accommodate the placement of
the robot, as described above, plus an external flat screen computer monitor,
keyboard and mouse pad in such a manner that they can be readily viewed and
accessed. Shelving shall be provided below the table to allow the placement of
a desk top computer tower (~7 x 20 x 17 in.), two reservoir bottles, a vaccum
pump and vacuum trap apparatus (~ 2 x 2 x 2 ft) and provisions shall be made for
the connection of an electrical cord from the computer and tubing from the
reservoir bottles and vacuum trap to the robot. Provisions must also be made
for the installation of a waste receptacle below the back, left side of the
robot platform for collection of used pipet tips that are dispensed from the
robot. Provisions shall be made for service personnel to gain access to the
right side of the robot (technical tower) and the top of the robotic arm for
maintenance or repairs, and/or it shall be possible to remove the table from the
enclosure. The enclosure/table components shall be preassembled or assembled on
site by the vendor. The vendor shall warrant that the enclosure shall pass a
HEPA filter leak test (NSF/ANSI 49-2004 section A.3) and airflow smoke pattern
test (inward flow test only, NSF/ANSI 49-2004 section A.11) performed by an
authorized EPA-contracted certification company. All Items shall be delivered
to the U.S. EPA, 26 W. MLK Drive, Cincinnati, OH 45268. F.O.B. point shall be
destination. Price quotes shall include any shipping charges. The provision at
FAR 52.212-1, Instructions to Offerors-Commercial, applies to this to this
acquisition. The provision at FAR 52.212-2, Evaluation-Commercial Items is
applicable and the specific evaluation criteria are as follows: (1)Technical
acceptability of the item offered to meet the Government's requirement and (2)
price. Technical acceptability shall be evaluated on a pass or fail basis.
Offers shall provide descriptive technical literature in sufficient detail to
demonstrate that the items offered meet the minimum requirements specified
above. Offerors shall include a completed copy of the provision at FAR 52.212-3,
Offeror Representations and Certifications-Commercial Items, with its offer.
This clause may be found on the Internet at the following site:
http://www.arnet.gov/far/. Clause FAR 52.212-4, Contract terms and
Conditions-Commercial Items, applies to this acquisition. The Clause at FAR
52.212-5, Contract Terms and Conditions Required to Implement Statutes or
Executive Orders-Commercial Items, applies to this acquisition. Under FAR
52.215-5 paragraph (b), the following additional clauses apply to this
acquisition: FAR 52.222-21, Prohibition of Segregated Facilities; FAR 52.222-22,
Previous Contracts and Compliance Reports; FAR 52.222-25, Affirmative Action
Compliance; FAR 52.222-26 Equal Opportunity; FAR 52.222-35, Affirmative Action
for Disabled Veterans and Veterans of the Vietnam Era; FAR 52.222-36
Affirmative Action for Workers with Disabilities; FAR 52.222-37 Employment
Reports on Disabled Veterans and Veterans of the Vietnam Era; FAR 52.225-1, Buy
American Act--Balance of Payments Program - Supplies; FAR 52.225-13,
Restrictions on Certain Foreign Purchases; FAR 52.225-15, Sanctioned European
Union Country End Products; FAR 52.204-7 Central Contractor Registrations, and
FAR 52.232-33, Payment by Electronic Funds Transfer - Central Contractor
Registration. Offerors shall submit 1 copy of their quotation, that references
RFQ-OH-05-00028, no later than TUESDAY, JANUARY 4, 2005 at 10:00 AM EST to Mr.
Josh Bowers as follows: via email at [Please use the link below], or via fax at
513-487-2109. Questions or comments may be directed to Mr. Josh Bowers at
513-487-2104 or via e-mail at [Please use the link below].


Company Reports

Get the Dun & Bradstreet Report for Environmental Protection Agency, Office of Acquisi
Company Information
 
Company Name:   Environmental Protection Agency, Office of Acquisi
Address:Login to view contact details
City:Login to view contact details
State/Province:Login to view contact details
Country:Login to view contact details
Zip/Postal Code:Login to view contact details
Phone:Login to view contact details
Fax:Login to view contact details
Contact Person: Login to view contact details
Email:Click here to send this company an email

Didn't find what you were looking for? Post a lead!

Click here to forward this trade lead to a friend!


Worldbid.com Affiliates